9 TROUBLE-SHOOTING GUIDE

Problem

Cause

Solution

No
display
when
turned on

a) Batteries not in place
b) Batteries not in

correct polarity (+
and – position).

c) Weak batteries

a. Remove rubber

boot/stand. Check
that batteries are in
place and making
good contact.

b) Reinsert batteries

with correct polarity.

c) Replace batteries.

Unstable
readings

a) Insufficient

electrolyte in probe.

b) Air bubbles trapped

around the probe

c) Dirty or damaged

probe

d) Probe not deep

enough in sample

e) External noise

pickup or induction
caused by nearby
electric motor.

f) Broken probe.

a) Fill probe with

electrolyte & replace
pre-membraned cap

b) Stir or tap probe to

remove bubbles.

c) Clean the probe and

re-calibrate

d) Make sure sample

entirely covers the
probe sensors.

e) Move or switch off

interfering motor.

f) Replace probe.

Slow
response

a) Dirty / Oily probe.
b) Temperature is

changing.

a) Clean probe. See

“Probe Care &
Maintenance”.

b) Allow time for

temperature to
stabilize

Not
respondin
g to key
press

a) HOLD operation

indicated by flashing
display.

b) Damaged key pad
c) Internal program

error.

a) Cancel HOLD mode.
b) Return to dealer.
c) Reset all internal

programs by
reinserting batteries.

42