LaMotte TDS 6 Conductivity User Manual
Page 8
Keypad
The CON 6 /TDS 6 meter has 6 buttons on the splash-proof keypad; ON/OFF,
HOLD/ENTER, CAL, MODE,
and
question buttons. Some buttons have
several functions depending on the mode of operation.
Powers on and shuts off the meter. Goes directly
into the measurement mode when the meter is
turned on.
Enters into the calibration mode for
Conductivity/TDS and Temperature.
To abort calibration or set-up mode without
confirming any set value.
HOLD: Freezes the measured reading. To activate,
press HOLD button while in measurement mode. To
release, press HOLD button again.
ENTER: Press to confirm values in the calibration
mode, and to confirm selections in the SET-UP
mode.
In Calibration Mode: Press to scroll through
calibration values.
In Set-up Mode: Press to scroll through the set-up
sub-group programs.
Press
button during conductivity measurement
mode to activate manual ranging function. Each
button press will move to a higher conductivity
range.
Selects measurement mode for Conductivity/TDS
and Temperature.
Pressed together with ON/OFF button to go to the
SET-UP mode. This allows customization of meter
preferences such as selecting the electrode cell
constant, normalization temperature, temperature
coefficient factor, TDS factor (for TDS 6), automatic
calibration (CON 6 only) or manual calibration,
single-point or multi-point calibrations, and to reset
meter to factory default.
ON
OFF
CAL
HOLD
ENTER
MODE
8