LaMotte TDS 6 Conductivity User Manual
Page 27
Temperature Coefficient
The temperature coefficient is the amount of change in
conductivity per degree of temperature; it is expressed in
percent per °C. Entering the exact temperature
coefficient of the solution being measured will accurately
compensate the temperature for almost any solution. The
temperature coefficient can be set between 0.0 and 3.0 %
per °C. The meter default is 2.1% per °C.
6. Enter Advanced Set-up.
7. Press
or
button until [t.Co %] appears on the
LCD. Press ENTER button.
8. Press
or
button to select a value between 0.0 and
3.0.
9. Press ENTER button to select. The meter will return
to the menu, [t.Co %].
Press
or
button to move to the next menu or press CAL to exit to the
measurement mode.
Normalization Temperature
The meter can be set to normalize the conductivity measurements to a
standard temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1. Enter Advanced Set-up.
2. Press
or
button until [t.nr °C] appears on the
LCD. Press ENTER button.
3. Press
or
button to select either [25.0 °C] or [20.0
°C].
4. Press ENTER button to select. The meter will return
to the menu, [t.nr °C].
5. Press
or
button to move to the next menu or press
CAL to exit to the measurement mode.