LaMotte TDS 6 Conductivity User Manual
Page 22
NOTE:
If the value of the solution being measured is higher than the range selected
[Or] will appear on the primary display. Press RANGE until the correct range
is selected.
The meter resets to the Auto-ranging function once it is turned off. The
manual ranging function must be reset each time the meter is turned on.
HOLD Function
The hold feature will freeze the display for a delayed
observation. HOLD can be used any time in
measurement mode.
1. To hold a measurement, press the HOLD button
while in measurement mode. [HO] will appear on the
display.
2. To release the held value, press the HOLD button
again. Continue to take measurements.
NOTE:
The meter will shut off automatically after 20 minutes of nonuse.
If the meter is shut off either automatically or manually, the HOLD value will
be lost.
S
µ
mS
mS
Auto-ranging
Manual ranging: 0 - 20.00 uS/cm
Manual ranging: 0 - 200.0 uS/cm
Manual ranging: 0 - 2000 uS/cm
Manual ranging: 0 - 20.00 mS/cm
Manual ranging: 0 - 200.0 mS/cm
S
µ
S
µ
S
µ
S
µ
S
µ
HOLD
ENTER
HO
22