beautypg.com

LaMotte Tracer Total Chlorine Meter User Manual

Page 14

background image

TROUBLESHOOTING
Problem

Check

Action

Power on but no
display

Batteries

Insert batteries

Verify correct polarity

Replace

Unstable reading

Electrode

Immerse probe more deeply in sample

Remove air bubbles caught under
electrode

Clean electrode

Replace electrode

Slow response

Electrode

Clean electrode

Unexpected low
results

Tablet

Confirm tablet has been added

Unexpected high
results

Tablet

Dissolve tablet completely

Display frozen

HOLD
function

Press MODE/HOLD or turn meter
off. Restart.

Button
press

Remove batteries (stored data will be
lost)

EXPANDING YOUR TRACER
Interchangeable electrodes are available to convert the Total Chlorine
TRACER to a pH TRACER or an ORP TRACER.
Remember to ask for instructions and appropriate reagent or buffer
tablets when ordering pH or ORP electrodes.

pH TRACER Electrode, 0.00-14.00 pH

Code 1733

The pH TRACER Electrode is used with pH 4.0, 7.0
and 10.0 buffers. Order using the following codes:

pH 4.0 Mini Buffer Tablets (100)

Code 3983A-J

pH 7.0 Mini Buffer Tablets (100)

Code 3984A-J

pH 10.0 Mini Buffer Tablets (100)

Code 3985A-J

14