beautypg.com

LaMotte Tracer Total Chlorine Meter User Manual

Page 12

background image

MAINTENANCE
Storage
When not in use, the TRACER should be stored dry with the electrode
cap in place. Cover the connector with the clear cover whenever the
electrode is removed from the meter body.
Battery Replacement
Battery Replacement
1. Twist off the battery compartment cap.
2. Hold the battery housing in place with one finger. Remove the

battery carrier by pulling on the small tabs.

3. Replace the four CR2032 batteries. Observe polarity.
4. Replace the battery compartment cap.

Cleaning the Electrode
The TRACER Chlorine Electrode arrives in a factory-cleaned condition
and is ready to be used. The electrode will perform better and yield more
stable readings after the first few tests. Firmly wipe the electrode surface
with a clean, dry paper towel before and after use to remove any
compounds that may have formed. Rinse the electrode in deionized water
to clean it. When not in use, the TRACER should be stored with the
electrode cap in place.

12