beautypg.com

LaMotte Tracer Total Chlorine Meter User Manual

Page 13

background image

Replacing the Electrode
The TRACER is shipped with an electrode attached. If the electrode
needs to be replaced or changed, follow these steps.
1. Press the ON/OFF button to turn the meter off.
2. To remove an electrode, unscrew and completely remove the

electrode collar. Turn the collar counter-clockwise.

3. Gently rock the electrode from side to side, pulling it downwards,

until it disconnects from the meter.

4. To attach an electrode, align the slots and carefully plug the electrode

into the meter socket.

5. Firmly tighten the electrode collar to create a seal with the rubber

gasket between the electrode and the meter. Do not over tighten.

6. Cover the connector end of the removed electrode with the small

clear plastic cap.

13

.

12