HF scientific Micro200BW Turbidimeter - 0 -1000 NTU User Manual
Page 29

Page 15
MICRO 200 BW 0 - 1000 NTU (5/05)
Rev. 3.1
5. Display Parameters
This is the Display parameters Menu. The sub-menu selections are described below. In the example above:
Upper Limit is set to 999 NTU
Lower Limit is set to zero NTU
Time Base is set to 60 minutes
D to A is set for 4 to 20 mA
Note: The D to A automatically calibrates to upper & lower (for our example you will get 4mA at 0 NTU & 20 mA @ 999
NTU)
Selection Options:
F1:
Selects UPPER LIMIT MENU
F2:
Selects LOWER LIMIT MENU
F3:
Selects Time Base between 60 minutes and 24 hours
F4:
Selects Set Average Menu
F5:
Selects D to A between 4-20 mA or 0 to10 volts
a. Set Upper Limit
This is the Set Upper Limit Menu. This is where you can change the upper limit for both the graphical display and the high setting
to which the D to A output is calibrated. The set lower limit menu is identical to this and is not shown.
Selection Options:
F1:
Not Used
F2:
Selects the Digit to change as indicated by a flashing cursor
F3:
Increments the selected Digit (increases it)
F4:
Decrements the selected digit (decreases it)
F5:
Not Used
F2
F3
ALARM
F5
F4
MENU
F1
Display Parameters
Upper Limit
999.
Lower Limit
ØØØ.
Select Time Base
6Ø Min.
Set Average
Select D to A
4 to 2Ø mA
F2
F3
F4
F1
Set Upper Limit
Flashing cursor
ALARM
F5
MENU
999.
Digit
Up
Down