10.4 Diagnostic Chart
Symptom
Cause Cure
Lower display shows
MA
4-20 mA loop open
Check wiring. See sections 3.3.4
and 7.2
Lower display shows
DESC
Desiccant
pouch
bad
Change desiccant pouch. See
section 9.2
Lower display shows
LAMP
Lamp failed
Replace lamp. Refer to section 9.3
Lower display shows
CLN
Ultrasonic cleaning failure
Refer to section 8.1
Lower display shows
CAL Err
after a calibration.
100%T Calibration failure
Refer to section 5.2
Lower display shows
FAIL
Major system fault
Refer to section 10.1 & 10.2
Readings are lower than expected
(1) Bubbles in solution
(2) Condensate or leaky
cuvette
(3) Flow through cuvette
dirty
Instrument out of calibration
Increase P-ST. Refer to section 7.13
Check flow through cuvette for
condensate or leaks.
Clean cuvette. See section 9.1
Recalibrate. Refer to section 5
Readings are erratic
(1) Bubbles in solution
(2) Debris in flow through
(1) See above
(2) Clean debris from cuvette
Readings are higher than expected
Instrument out of calibration
Recalibrate. Refer to section 5
10.5 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do not hesitate to
contact either the HF scientific Technical Service Department or the HF scientific
Customer Service Department. Contact information is shown on section 10.2.