9.8 System Cleaning
The suggested cleaning period is once every three months. As this is very site specific, this
period will have to be adjusted as required.

Cleaning is recommended with chlorine to remove algae and then a Lime-A-Way to
remove calcium and lime. The system should be flushed after cleaning with each solution.
Steps:

1. Press the

MODE/EXIT key to change the instrument to the SVC mode. This will

ensure the pump will not start.

2. Disconnect the main drain and inlet tubings.
3. Install shorter tubings to perform the cleaning locally. These should be just long

enough to reach a bucket placed under or near the system.

4. Clean the T-Strainer. As described in section 9.5 above.
5. Drop both the source line and the drain line into a bucket containing 6% Sodium

Hypo chloride (bleach) solution.

6. Initiate a pump cycle by pressing the

MODE/EXIT key to switch to the pre-

AUTO mode, and then press

to return to

AUTO and run the pump. Allow the

pump to run through its full pumping cycle.

7. Press the

MODE/EXIT key to change the instrument to the SVC mode. then

transfer the source line into a container at least 8 liters (2 gallons) of clean water.
Route the drain either to another container or to a convenient drain.

8. Initiate a pump cycle by pressing the

MODE/EXIT key to switch to the pre-

AUTO mode, and then press

to return to

AUTO and run the pump. Allow the

pump to operate until the container is empty.

9. Press the

MODE/EXIT key to change the instrument to the SVC mode.

10. Drop both the source line and the drain line into a bucket containing a solution of

50% Lime-A-Way and 50% water.

11. Initiate a pump cycle by pressing the

MODE/EXIT key to switch to the pre-

AUTO mode, and then press

to return to

AUTO and run the pump Allow the

pump to operate through its full pumping cycle.

12. Flush by repeating steps 7-9 above.
13. Reconnect the normal inlet and main drain lines.
14. Switch to

AUTO mode to continue normal operation.