7.15 Desiccant Alarm
When the humidity detector in the AccUView Wastewater indicates that the internal
environment is close to the point where humidity could cause condensation, the instrument
will display
DESC as a warning. If desired, a desiccant warning can activate the alarm and
send the 4-20mA to 2mA. To activate the alarm when the desiccant fails, select
On in the
DESC menu. The default for this menu is OFF. Make selections using the
t and u
buttons then press the
↵ button to move to return to pre-AUTO mode. Then press the ↵
button to return to normal operation.
7.16 Saving Configuration Settings
If extended settings is set to
OFF, pressing the
↵
button will save all settings and the
AccUView Wastewater will be sent to
pre-AUTO mode. Press the ↵ button to return to
normal operation.
If extended settings is set to
On, after the last menu of the extended settings, pressing the
↵
button will save all settings and the AccUView Wastewater will be sent to
pre-AUTO
mode. Press the
↵ button to return to normal operation.
The
CONFIG menu may be used at any time to reset or change any of the parameters.
The
CONFIG menu may be exited at any point in the menu by using the MODE/EXIT
key. Any features that have been modified will be saved.
AccUView Wastewater (1/09)
23