beautypg.com

15 desiccant alarm, 16 saving configuration settings – HF scientific AccUView Wastewater UV %Transmission Monitor User Manual

Page 29

background image

7.15 Desiccant Alarm
When the humidity detector in the AccUView Wastewater indicates that the internal
environment is close to the point where humidity could cause condensation, the instrument
will display

DESC as a warning. If desired, a desiccant warning can activate the alarm and

send the 4-20mA to 2mA. To activate the alarm when the desiccant fails, select

On in the

DESC menu. The default for this menu is OFF. Make selections using the

t and u

buttons then press the

↵ button to move to return to pre-AUTO mode. Then press the ↵

button to return to normal operation.










7.16 Saving Configuration Settings
If extended settings is set to

OFF, pressing the

button will save all settings and the

AccUView Wastewater will be sent to

pre-AUTO mode. Press the ↵ button to return to

normal operation.

If extended settings is set to

On, after the last menu of the extended settings, pressing the

button will save all settings and the AccUView Wastewater will be sent to

pre-AUTO

mode. Press the

button to return to normal operation.

The

CONFIG menu may be used at any time to reset or change any of the parameters.

The

CONFIG menu may be exited at any point in the menu by using the MODE/EXIT

key. Any features that have been modified will be saved.

AccUView Wastewater (1/09)

23

Rev. 2.9