Metrohm 762 IC Interface User Manual
Page 35

4.1 Technical data
762 IC Interface
31
Zero error
±
2.5 mV
Pin assignment
Pin 1,4,5,8,9: GND (ground, connected to
housing)
Pin 2:
Analog 1+
Pin 3:
Analog 1–
Pin 6:
Analog 2+
Pin 7:
Analog 2–
Safety specifications
Construction/testing
According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference
Standards met:
EN55011 (level B), EN55022 (level B),
EN 50081-1/2, EN61000-3-2
Immunity to interference
Standards met:
IEC61000-4-2/EN61000-4-2 (level 3),
IEC61000-4-3/EN61000-4-3 (level 3),
IEC61000-4-4/EN61000-4-4 (level 4),
IEC61000-4-5/EN61000-4-5 (level 2/3),
IEC61000-4-6/EN61000-4-6 (level 3),
IEC61000-4-11/EN61000-4-11,
IEC61000-4-14/EN61000-4-14 (level 3),
EN50082-2, NAMUR
Ambient temperature
Nominal operating range
+5
…
+45°C
(at 20
…
80 % atmospheric humidity)
Storage, transport
–40
…
+70°C
Housing
Material of cover
Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, CFC-free
Material of base
Steel, enameled
Dimensions
Width
255 mm
Height
128 mm
Depth
340 mm
Weight (with accessories)
2.762.0010: 4.3 kg
2.762.0020: 4.9 kg