3 Operation
762 IC Interface
26
Program
On the
Program
subpage, program steps including time, program in-
struction and parameter can be entered.
First column
Time at which program instruction is applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is
applied together with the last instruction with
time entry.
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the
Events configuration
tab.
Third column
Parameter for program instruction (see below).
ENABLED
Enable program start (a disabled program is
not started).
<
Add
>
Add new program instruction.
<
Delete
>
Delete selected program instruction.
<
Verify
>
Test the time program (error messages are dis-
played if program is wrong).
List of program instructions
The following program instructions can be added to the time program
on the
Program
subpage:
Instruction
Entry
Meaning
Events
0, 1, p, *
Set
event output lines 1...7 to the desired
values. For entry of the first value, enter
1
,
0
,
p
or
*
. For entry of the other values, move
the cursor in front of the value to be
changed and enter
1
,
0
,
p
or
*
.
Program
END, RESET
The
END
flag can be used to end a
program, especially if the program time
should be longer than the chromatogram
duration. Additional steps after this flag are
not allowed. The
RESET
flag is used to
reset the parameters to the system startup
values.