2 mains connection – Metrohm 762 IC Interface User Manual
Page 14

2 Installation
762 IC Interface
10
2.2
Mains connection
2.2.1
Mains voltage and fuses
The 762 Interface has a power supply which automatically adjusts itself
to the existing mains voltage (100
…
240 V) and frequency (50
…
60 Hz).
It is equipped with an electronic overload protection device and also
has two fuses; however, these should only be exchanged by Metrohm
service technicians.
2.2.2
Mains cable
The instrument is supplied with one of three mains cables
•
6.2122.020 with plug SEV 12 (Switzerland,
…
)
•
6.2122.040 with plug CEE(7), VII (Germany,
…
)
•
6.2133.070 with plug NEMA 5-15 (USA,
…
)
which are three-cored and fitted with a plug with an earthing pin. If a
different plug has to be fitted, the yellow/green lead (IEC standard)
must be connected to the earthing socket 77 (protection class 1).
Any break in the earthing inside or outside the instrument can make it
a hazard!
Plug the mains cable into mains connection plug 88 at the 762 IC Inter-
face (see Fig. 3 and Fig. 5).
2.2.3
Switching the instrument on/off
The 762 IC Interface is switched on and off using mains switch 11 (see
Fig. 2 and Fig. 4). When the instrument is switched on the mains pilot
lamp 22 lights up.