8 technical specifications, 1 reference conditions, 2 uv/vis detector – Metrohm 887 Professional UV/VIS Detector User Manual
Page 45: Reference conditions, Uv/vis detector

■■■■■■■■■■■■■■■■■■■■■■
8 Technical specifications
887 Professional UV/VIS Detector
■■■■■■■■
37
8 Technical specifications
8.1
Reference conditions
The technical specifications listed in this chapter refer to the following ref-
erence conditions:
Ambient tempera-
ture
+25 °C (± 3 °C)
Instrument status
> 30 min in operation with both lamps switched on.
8.2
UV/VIS detector
Measuring range
Absorbtion
–2.0…+2.0 AU
Number of
channels
8 measuring channels, 4 of them readable in analog mode
Resolution
2.5 x 10
-7
AU
Noise
2.5 x 10
-5
AU (at data rate 1/s)
Drift
at reference
conditions
5 x 10
-4
AU/h
Wavelength range
wavelength
range
λ
190…900 nm
Wavelength
bunching
Δλ
1…limit nm (max. 101)
limit is 2 times the maximum number nm between the wavelength of
the corresponding channel and the nearest wavelength border.
Δλ is
always odd, since symmetrical around the wavelength of the corre-
sponding channel including the mean value.
Accuracy abso-
lute
± 3 nm
Stability
± 1 nm (over temperature range)
Optical resolu-
tion
5 nm (at 254 nm)
Measuring interval
Data rate for
each channel
0.5…50 Samples/sec