6.3 Replacing the UV lamp

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

887 Professional UV/VIS Detector

Removing old UV lamp

Warning

After longer operation, the UV lamp becomes hot!

Risk of burns!

Allow the lamp to cool down, before you begin.

Switch off the instrument, before replacing the lamp.

1

2

3

1 Disconnecting the UV lamp

Loosen the retaining ring of the UV lamp plug and pull the plug out
of the UV connection socket.

2 Loosening the fastening ring

Loosen and remove the fastening ring of the UV lamp.

3 Removing UV lamp

Carefully remove the old UV lamp from its housing.