6 Handling and maintenance

887 Professional UV/VIS Detector

■■■■■■■■

33

If the noise persists, the flow-through cell must be dismantled and
cleaned manually (see "Dismantling and cleaning the flow-through
cell", page 33)
.

Dismantling and cleaning the flow-through cell

Note

Ex works, the retaining screws are tightened with a torque wrench. This
guarantees a pressure stability up to 5MPa (50bar).

Once the flow-through cell has been opened and closed again, this
maximum pressure stability can no longer be guaranteed.

Preconditions:

Rinsing the flow-through cell with solvent was inefficient.

Required tools:

To open the measuring cell, you need a slotted screwdriver size 5.

Seals from seal set 6.2764.000

1 Removing the flow-through cell

Disconnect inlet capillary and outlet capillary,

Loosen the knurled screws and remove them,

Remove the flow-through cell from the cell holder.

5

4

3

1

4

3

1

2

6

Figure 13

Flow-through cell – parts