Metrohm 887 Professional UV/VIS Detector User Manual
Page 42

6.6 Cleaning the flow-through cell
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
887 Professional UV/VIS Detector
1
Retaining screw
2
Outer seal
3
Lens
4
Inner seal
5
Measuring cell – opening
6
Cell holder
2 Removing the lens
■
Loosen the retaining screw (13-1) with a slotted screwdriver size
5 and remove it.
Remove the seal from the inside of the retaining screw (13-2).
■
Carefully remove the lens (13-3) and the inner seal (13-4) from
the measuring cell.
■
Put the lens onto a white paper, this allows you to see possible
deposits.
3 Cleaning the lens
■
Clean the lens with ultrapure water, methanol or another suitable
solvent (depending on the application) and wipe with a lint-free
cloth.
■
Rinse again with ultrapure water and dry with a lint-free cloth.
4 Re-inserting the lens
■
Insert a new inner seal (13-4) into the measuring cell, make sure
that it lies flat and centered in the recess of the measuring cell.
■
Insert the lens (13-3) with its flat side in into the measuring cell,
make sure that it lies flat an centered in the recess of the measur-
ing cell.
■
Insert an new outer seal (13-2) into the retaining screw.
■
Insert the retaining screw (13-1) and tighten it with a slotted
screwdriver size 5.
5 Cleaning the second lens
Repeat steps 2 – 4 at the opposite side of the cell holder.
6 Re-inserting the flow-through cell
Follow steps 3 – 4 of "Installing the flow-through cell", page 18.