beautypg.com

Metrohm 887 Professional UV/VIS Detector User Manual

Page 42

background image

6.6 Cleaning the flow-through cell

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

887 Professional UV/VIS Detector

1

Retaining screw

2

Outer seal

3

Lens

4

Inner seal

5

Measuring cell – opening

6

Cell holder

2 Removing the lens

Loosen the retaining screw (13-1) with a slotted screwdriver size
5 and remove it.
Remove the seal from the inside of the retaining screw (13-2).

Carefully remove the lens (13-3) and the inner seal (13-4) from
the measuring cell.

Put the lens onto a white paper, this allows you to see possible
deposits.

3 Cleaning the lens

Clean the lens with ultrapure water, methanol or another suitable
solvent (depending on the application) and wipe with a lint-free
cloth.

Rinse again with ultrapure water and dry with a lint-free cloth.

4 Re-inserting the lens

Insert a new inner seal (13-4) into the measuring cell, make sure
that it lies flat and centered in the recess of the measuring cell.

Insert the lens (13-3) with its flat side in into the measuring cell,
make sure that it lies flat an centered in the recess of the measur-
ing cell.

Insert an new outer seal (13-2) into the retaining screw.

Insert the retaining screw (13-1) and tighten it with a slotted
screwdriver size 5.

5 Cleaning the second lens

Repeat steps 2 – 4 at the opposite side of the cell holder.

6 Re-inserting the flow-through cell

Follow steps 3 – 4 of "Installing the flow-through cell", page 18.