8 configuration of the export template – Metrohm TitrIC 4 User Manual
Page 13

Application Bulletin 318 e
Installation Instructions of TitrIC 4
Page 13/16
TitrIC 4 - preparation of titration equipment:
This method is used for preparing all titration instruments,
i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo
TM
– not from TitrIC!
TitrIC 4 - IC calibration:
This method is used for calibrating the IC measurement with standard
solutions without the titration part of a typical TitrIC measurement; this means that a lot of
standard solution can be saved.
TitrIC 4 – semi automated pH calibration:
With this method the pH electrode is calibrated semi-
automatically; the buffer solutions are flushed into the titration vessel. Start this method only
from tiamo
TM
– not from TitrIC!
TitrIC 4 - pH calibration:
Here the Aquatrode is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo
TM
!
TitrIC 4 - titer determination of EDTA:
This method is used for the titer determination. Start this
method only from tiamo
TM
!
TitrIC 4 - titer determination of HCl:
This method is used for the titer determination. Start this method
only from tiamo
TM
!
TitrIC Calibrate 856 conductivity module
This method is used for calibration of the conductivity
module. Start this method only from tiamo
TM
!
Import: Click on
Method
at the left, then select under
File / Method manager… / Edit / Import…
in the
following directory on the CD you will find all methods:
\Methods\Titration
.
Note!
Alterations of these methods should only be carried out by a person at the administrator level
and who is thoroughly familiar with tiamo
TM
. Never forget that in TitrIC this new method must
be selected in the sample table under
Methods tiamo
.
6.2.8 Configuration
of the export template
Export template: tiamo
TM
must transfer its results to TitrIC in a particular way, which means that the
export template must be generated exactly in the following form:
Under
Database / Tools/ Templates / Export templates / New
enter the following:
Configuration of the export template in TitrIC
The target directory for the report is as following:
Windows 2000, Windows XP:
C:\Dokumente und Einstellungen\All Users\Application Data\Titric\reports
Vista, Windows 7:
C:\ProgramData\Titric\reports
In Windows Vista and Windows 7 hidden folders need to be made visible in the explorer to get access.
Then click on
Select fields
and enter the following: