TitrIC 4 - preparation of titration equipment:

This method is used for preparing all titration instruments,

i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo

TM

– not from TitrIC!

ƒ

TitrIC 4 - IC calibration:

This method is used for calibrating the IC measurement with standard

solutions without the titration part of a typical TitrIC measurement; this means that a lot of
standard solution can be saved.

ƒ

TitrIC 4 – semi automated pH calibration:

With this method the pH electrode is calibrated semi-

automatically; the buffer solutions are flushed into the titration vessel. Start this method only
from tiamo

TM

– not from TitrIC!

ƒ

TitrIC 4 - pH calibration:

Here the Aquatrode is taken out of the titration vessel and placed into the

buffer solutions. Start this method only from tiamo

TM

!

ƒ

TitrIC 4 - titer determination of EDTA:

This method is used for the titer determination. Start this

method only from tiamo

TM

!

ƒ

TitrIC 4 - titer determination of HCl:

This method is used for the titer determination. Start this method

only from tiamo

TM

!

ƒ

TitrIC Calibrate 856 conductivity module

This method is used for calibration of the conductivity

module. Start this method only from tiamo

TM

!


Import:
Click on

Method

at the left, then select under

File / Method manager… / Edit / Import…

in the

following directory on the CD you will find all methods:

\Methods\Titration

.

Note!
Alterations of these methods should only be carried out by a person at the administrator level
and who is thoroughly familiar with tiamo

TM

. Never forget that in TitrIC this new method must

be selected in the sample table under

Methods tiamo

.

6.2.8 Configuration

of the export template

Export template: tiamo

TM

must transfer its results to TitrIC in a particular way, which means that the

export template must be generated exactly in the following form:

Under

Database / Tools/ Templates / Export templates / New

enter the following:

Configuration of the export template in TitrIC

The target directory for the report is as following:
Windows 2000, Windows XP:

C:\Dokumente und Einstellungen\All Users\Application Data\Titric\reports

Vista, Windows 7:

C:\ProgramData\Titric\reports

In Windows Vista and Windows 7 hidden folders need to be made visible in the explorer to get access.

Then click on

Select fields

and enter the following: