Application Bulletin 318 e
Installation Instructions of TitrIC 4
Page 10/16

5 Miscellaneous

5.1 Stirrer

The 802 Rod Stirrer is inserted through opening 9 of the external cell (see page 7) and connected to the
tower of the 815. The propeller (6.1909.020) is pushed on from below.

Length of the aspiration tips in the external cell

ƒ Clamp the titration vessel (6.1415.220) to the titration vessel lid (6.1414.060).

ƒ The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is

the only way to ensure that the external cell is cleaned properly.

ƒ The buret tips of the 3-way stopper in opening 6 and the buret tip of opening 3 should reach the

middle of the titration vessel. The three tubings of the 3-way stopper have to be mounted with
antidiffusion micro valves to ensure a precise titration.

5.2 Electrodes

ƒ The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 4. The

cable is connected to the Titrando at

Input 2

. The plug of the gray cable is connected to the

opening marked

Ind.

. The red plugs with the adapters 6.2103.140 and 6.2103.130 are inserted

in the openings marked

Temp.

If the electrode is not used for some time then it should be returned to its holder and kept in the
storage solution (6.2323.000).

ƒ The Ca-ISE (Polymer membrane electrode Ca, 6.0508.110) is connected to the electrode cable

(6.2104.020) and inserted into opening 8. The cable is connected to the Titrando at

Input 1

. The

plug of the gray cable is connected to the opening marked

Ind.

.

ƒ To have a reference electrode for the Ca-ISE one can short-circuit the reference channel of the

Ca-ISE with the one of the Aquatrode. Thereto connect

Ref.

of

Input 1

with

Ref.

of

Input 2

with the

electrode cable 2x plug B (6.2106.020).

ƒ Conductivity cell (6.0915.100): The measuring cell cable is led through the guide chain of the

815 and the plug is inserted into the opening marked

Cond. Cell

of the 856 Conductometer. Later

(as soon as the titration head has been mounted) the measuring cell with the SGJ sleeve
(6.1236.050) will be inserted in the SGJ 14 opening of the titration head.

6 Software

6.1 Installation of the software

ƒ tiamo

TM

2.1 – installation (tiamo

TM

CD): All the standard directories proposed by the program

should be accepted.

ƒ MagIC Net 2.0 – installation (MagIC Net CD): All the standard directories proposed by the

program should be accepted. Restart windows.

ƒ TitrIC 2.0 – installation (TitrIC CD): TitrIC 2.0 must only be installed after tiamo

TM

2.1 and

MagIC Net 2.0 have already been installed. After installation the standard TitrIC directory can
be found under

C:\Program files\ Metrohm\TitrIC

, the TitrIC.exe file under

C:\Program

files\Metrohm\tiamo\bin

. The TitrIC database and reports are found on

C:\Documents and Settings\All

Users\ApplicationData\TitrIC.

Æ All programs should be closed!