4 Operation and maintenance

IC equipment: Inline-Ultrafiltration

■■■■■■■■

21

Tighten the PTFE capillary (6.1803.060) to the inlet labeled IN of
the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).

Tighten the PTFE capillary (6.1803.080) to the outlet labeled OUT
of the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).

Tighten the PTFE capillary (6.1803.050) to the outlet labeled OUT
of the ultrafiltration cell's upper chamber using a PVDF pressure
screw (6.2744.000).

4 Inserting the ultrafiltration cell into the holder

Place the ultrafiltration cell back in the holder (see Chapter 3.3,
page 14)
.

5 Deaerating the ultrafiltration cell

Carry out all steps contained in the instructions Rinsing the ultrafiltra-