3.2 Connecting the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
IC equipment: Inline-Ultrafiltration
■
Pump tubing LFL (orange/yellow), 3 stoppers (6.1826.320)
■
Pump tubing LFL (yellow/yellow), 3 stoppers (6.1826.390)
■
PEEK capillary, 0.5 mm ID / 1 m (6.1831.060)
■
PEEK capillary, 0.5 mm ID / 70 cm (6.1831.160)
■
Pressure screw PVDF (6.2744.000)
■
Coupling olive/UNF 10/32 (6.2744.030)
■
Pressure screw, short (6.2744.070)
■
Pump tubing connection with locking nut (6.2744.160)
1 Preparing the pump tubing for the sample
Use the pump tubing with yellow stoppers (6.1826.390) for convey-
ing the sample.
■
Attach the coupling olive/UNF 10/32 (6.2744.030) to the inlet.
■
Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor).
2 Preparing the pump tubing for the filtrate
Use the pump tubing with orange/yellow stoppers (6.1826.320) for
conveying the filtrate.
■
Attach the coupling olive/UNF 10/32 (6.2744.030) to the inlet.
■
Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor).
3 Connecting the capillaries to the pump tubing for the sample
■
Tighten the PEEK capillary (6.1831.160) to the inlet of the pump
tubing with yellow stoppers (6.1826.390) using a pressure screw
(6.2744.070).
■
Tighten the PTFE capillary (6.1803.060) to the outlet of the pump
tubing with yellow stoppers (6.1826.390) using a pressure screw
(6.2744.070).