beautypg.com

Metrohm Inline-Ultrafiltration User Manual

Page 16

background image

3.1 Preparing the ultrafiltration cell

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

IC equipment: Inline-Ultrafiltration

7 Assembling the ultrafiltration cell

Place the upper part of the ultrafiltration cell on the lower part in
such a way that the two guide bolts fit exactly into the two bore
holes.

8 Screwing the ultrafiltration cell together

Screw the five screws with the washers in the ultrafiltration cell by
hand first.

Then firmly tighten them with the hex key in crosswise sequence.