3.1 Preparing the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
IC equipment: Inline-Ultrafiltration
7 Assembling the ultrafiltration cell
■
Place the upper part of the ultrafiltration cell on the lower part in
such a way that the two guide bolts fit exactly into the two bore
holes.
8 Screwing the ultrafiltration cell together
■
Screw the five screws with the washers in the ultrafiltration cell by
hand first.
■
Then firmly tighten them with the hex key in crosswise sequence.