4.10 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
84
■■■■■■■■
850 Professional IC – Cation – Prep 2
Order
number
Name
Material
Inner
diameter
Use
6.1826.360
Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon)
1.25 mm
For inline sample dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon)
1.37 mm
For the sample solution in
inline ultrafiltration.
4.10.2.2
Pump tubing connection with filter
The 6.2821.130 filters (44-2) should be changed every 3 months, more
frequently at higher backpressure.
1
2
3
Figure 44
Pump tubing connection – Changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter housing
Replacing the filter
1 Unscrewing filter screw
■
Screw the filter screw (44-3) out of the tubing olive (44-1) with
the aid of two 6.2621.000 adjustable wrenches.
2 Replacing the filter
■
Remove the old filter (44-2) with tweezers.
■
Place the new filter (44-2) flat in the tubing olive (44-1) with
tweezers.
3 Mounting filter screw
■
Screw the filter screw (44-3) back into the tubing olive (44-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.