13 mains connection, 14 interfaces – Metrohm 850 Professional IC Cation - Prep 2 User Manual
Page 106
6.13 Mains connection
■■■■■■■■■■■■■■■■■■■■■■
96
■■■■■■■■
850 Professional IC – Cation – Prep 2
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)
Cell tempera-
ture
20…50 °C in increments of 5 °C
Temperature
stability
< 0.001 °C
Temperature
compensation
0…5 %/K adjustable, default 2.3 %/K
Heating up time
< 30 minutes (40 °C)
6.13
Mains connection
Required voltage
100…240 V ± 10 % (auto-sensing)
Required fre-
quency
50…60 Hz ± 3 (auto-sensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
internal fuse 3.15 A
6.14
Interfaces
USB
Input
1 USB upstream, type B (for connection to the PC)
Output
2 USB downstream, type A
MSB
2 MSB Mini Din 8-pin (female) (for Dosino, stirrer, remote lines, ...)
Caution
When connecting an instrument to the MSB connector you must
switch off the 850 Professional IC .
Detector
2 DSUB-15-pin high density (female)
Column recogni-
tion
3 (including 2 in the column thermostat (see Chapter 2.16, page 51))
Leak sensor