5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
90
■■■■■■■■
850 Professional IC – Cation – Prep 2
Problem
Cause
Remedy
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 4.13.4, page 87).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 60).
Conductivity detec-
tor is not recognized
in the software
No connection.
■
Check the cable connection (32-
■
Turn the instrument off and on again (after
15 seconds).
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Connections – dead vol-
ume in the system.
Check the connections (see Chapter 2.5, page
17) (use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).
Guard column – diminished
performance.
■
Replace guard column (see Chapter 2.19,
page 57).
Separation column – dead
volume at column head.
■
Install separation column in reverse flow
direction (if permitted by the information
sheet).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 60).
Precision problems -
significant scatter-
ing of the measured
values
Sample – Gas bubbles in
the sample.
Use sample degasser (see Chapter 2.13, page
43).
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 2.15.1, page 49).
Sample – Rinsing volume
too low.
Prolong the rinsing time (see Chapter 4.8,
page 80).
Injection valve – defective.
Request Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 4.13.4, page 87).