7 Technical specifications

940 Professional IC Vario ONE/HPG (2.940.1140)

■■■■■■■■

79

Gradient capabil-
ity

Isocratic or gradient (extendable to quaternary)

Profile

Step, linear, convex and concave

Resolution

< 1 nL/min

7.9

Injection valve

Actuator switch-
ing time

typ. 100 ms

Maximum operat-
ing pressure

35 MPa (350 bar)

Material

PEEK

7.10

Detector

You can find the technical specifications for the detector in the detector's
manual.

7.11

Sample degasser

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Time to establish
vacuum

< 60 s

7.12

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

Internal fuse 3.15 A