Metrohm 940 Professional IC Vario ONE/HPG User Manual

Page 45

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

940 Professional IC Vario ONE/HPG (2.940.1140)

■■■■■■■■

37

4 Deaerating the high-pressure pump

Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.

Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.

Open the purge valve using the rotary knob (approx. ½ turn).

Switch on the high-pressure pump in MagIC Net.

Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.

Switch off the high-pressure pump in MagIC Net.

Seal the purge valve using the rotary knob.

Remove the syringe from the purging needle.

Pull the purging needle out of the deaerating capillary.

5 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.