800 Dosino

7 Troubleshooting

7.1

Problems

Problem

Cause

Remedy

Data of the dosing
unit cannot be read.

Data chip of the dosing
unit mechanically dam-
aged or impaired by chem-
icals.

Remove the dosing drive and set it up
again.

Clean the data chip and the contact surfa-
ces.

Have the data chip replaced by the
Metrohm Service Dept.

Dosing unit recog-
nized either not at
all or incorrectly.

The dosing drive was not
attached correctly.

Remove the dosing drive and set it up
again.

Check whether the dosing drive is correctly
seated.

Switch the control instrument off and on
again.

If necessary contact Metrohm Service Dept.

Dosino becomes hot

Dosing drive is overloaded.
Valve disc or dosing piston
is blocked.

Switch off the instrument immediately.

Disassemble the dosing unit (see the man-
ual for the dosing unit) and clean all of the
individual parts. Replace defective parts.

Dosino cannot be
actuated by the con-
trol device.

Connection between Dos-
ino and control device is
interrupted or án error has
occurred on the Dosino.

Check the cable connections.

Switch the control instrument off and on
again.

Check the dosing and filling rate.

If necessary contact Metrohm Service Dept.

No dosing takes
place at all

Tubing connections are
blocked or dosing unit is
not assembled correctly.

Check whether the dosing tip is blocked.

Check whether the dosing port is sealed off
with a stopper.

Check whether the VENT port is sealed off
with a stopper (vacuum in the supply bot-
tle!). The VENT port must be open for pres-
sure compensation.

Remove the dosing drive and check
whether the dosing piston is connected to
the dosing drive. The piston stopper must
be flush with the upper side of the hous-
ing.