3 glp - validation, Glp - validation – Metrohm 800 Dosino User Manual
Page 39

■■■■■■■■■■■■■■■■■■■■■■
6 Handling and maintenance
800 Dosino
■■■■■■■■
31
6.3
GLP - Validation
Every drive and every dosing unit manufactured by the Metrohm Co. is
subjected to rigorous quality controls prior to shipment. Every dosing unit
is issued a quality certificate attesting conformance with the strict quality
criteria of the Metrohm Co. GLP( G ood L aboratory P ractice) requires,
among other things, periodic inspection of analytical measuring devices
with respect to precision and correctness on the basis of standard operat-
ing procedures S tandard O perating P rocedure, SOP ). This may also
include an inspection of dosing accuracy.
Recommended literature
■
Metrohm brochure "Quality management with Metrohm", detailed
information concerning the principles and procedural methods of
Good Laboratory Practice
■
Metrohm Applications Bulletin 283/1 "Validation of Metrohm burets"
The validation of burets is carried out by the Metrohm-Service with a spe-
cial software.
The Metrohm agents worldwide offer the possibility of on-site inspections
and certifications of dosing units and Dosinos with respect toaccuracy. It is
recommended that an accuracy inspection be performed when the dosing
cylinders and dosing pistons of a dosing unit are replaced.