2.4 Connections

Relay Box 731, Instructions for use

19

Control commands in Tinet 2.x:

In a sequence for 730 Sample Changer
Switch on aeration valve:

Switch off with bit pattern:

0*************

Switch 772 Pump Unit:

Bit pattern: *1************ (=on)
Bit pattern: *0************ (=off)

Switch Titrinos:

Bit pattern: **1*********** (=on)
Bit pattern: **0*********** (=off)

The output line can be selected freely. Avoid conflicts with other in-
struments which may be connected.

... – 731 Relay Box – 665, ... Dosimat

In an automated system 665, 715, 725, etc. Dosimats can also be
used as auxiliary dosing devices . If these cannot be connected di-
rectly to the control instrument (via dosing contacts) then this is pos-
sible with a 731 Relay Box via remote lines.

731

665

2x 6.2106.XXX cable strand, crossed