2. Installation

Relay Box 731, Instructions for use

6

Checking and replacing the AC fuses
Line voltage outputs AC1 and AC2 have their own fuses.

Fuse 5A(T)

AC 1

Use a screwdriver to press down strongly the affected fuse holder
and turn it counterclockwise through 90°. Carefully remove the
fuse holder.

Carefully remove the fuse from the fuse holder and check its
specifications:

220

250 V 5A (slow-blow)

Metrohm no. U.600.0023

Please note that the specification of this fuse is different from
that of the main fuse!

Replace the fuse if necessary and insert it in the fuse holder
again.

Fuse holder AC1