2.4 Connections

Relay Box 731, Instructions for use

11

2.4.2 The Metrohm remote socket

Particular instrument functions can be triggered by activating the in-
dividual pins (connection pins) of a remote socket. The instrument
itself activates certain pins according to the instrument’s condition
and transmits signal pulses when particular events occur.
The functions which are used for remote control in an automated
system are allocated to logical remote lines. They are subdivided
into input lines (signal inputs to trigger functions, e. g. START or
STOP) and output lines (signal outputs to display the instrument’s
condition or events, e. g. 'ready' or 'End of Determination' =EOD).

The pin assignation of a 25-pin remote plug can be seen from the
information given below.

Output 6
Output 7
Output 4
Output 2
Output 0
Output 8
Output 9
Output 10
Input 1
Input 3
Input 5
Input 7
Output 11

0 Vo lt

+5 Volt

Output 5
Output 3
Output 1

Output 12
Output 13

Input 0
Input 2
Input 4
Input 6

0 Vo lt

Remote plug

(with 25 pins)

+5V

active = low
inactive = high

active = low
inactive = high

Inputs

Inputs

Outputs

Outputs

Electrical switching of the inputs

and outputs of the remote socket

Standard allocation of the remote lines in Metrohm instruments:

Remote lines

Pin

Function

Input 0

21

Start

Input 1

9

Stop

Input 2

22

Enter

Input 3

10

Clear

Input 4

23

Sample Ready

Input 5

11

not used

Input 6

24

not used

Input 7

12

not used

Output 0

5

Ready

Output 1

18

Conditioning ok

Output 2

4

Determination busy

Output 3

17

EOD (End of Determination)

Output 4

3

L1

Output 5

16

Error

Output 6

1

Activate L3

Output 7

2

Pulse for recorder

Output lines 8…13 are normally not used as standard.