beautypg.com

Metrohm 944 Professional UV/VIS Detector Vario User Manual

Page 44

background image

8.2 UV/VIS detector

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

944 Professional UV/VIS Detector Vario (2.944.0010)

Integration
duration

10 - 247 ms

Duration of
measurement

50 - 2,000 ms

Analog output

Voltage range

–1.0 - +1.0 V

Resolution

30 µV

Noise

< 30 µV

Output impe-
dance

50

Ω (sustained short circuit-proof)

Offset error

±1.5 mV

Plug

9-pin DSUB plug (female)

1

2

3

4

5

6

7

8

9

0V

0V

0V

0V

0V

A4 (CH4) A3 (CH3) A2 (CH2) A1 (CH1)

Control input

Input voltage
range

0 - 5 V (5 V logic or switching contact control possible)

Impedance

22 k

Ω (permanent resistance up to 50 V. Secured against ESD.)

Function

Start, zero, 2x reserve
Caution: If "reserve 2" input is active when the instrument is switched
on, the instrument enters long-term test mode, which switches
through all elements following a time schedule and cannot be termina-
ted, not even by power failure. Can only be terminated by service soft-
ware or RS-232 commands.