beautypg.com

Metrohm 944 Professional UV/VIS Detector Vario User Manual

Page 40

background image

6.7 Cleaning the flow-through cell

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

944 Professional UV/VIS Detector Vario (2.944.0010)

2 Removing the lens

Loosen the retaining screw (12-1) with a slotted screwdriver, size
5, and remove it.
Remove the seal from the inside of the retaining screw (12-2).

Carefully remove the lens (12-3) and the inner seal (12-4) from
the measuring cell.

Put the lens onto a white paper, this allows you to see possible
deposits.

3 Cleaning the lens

Clean the lens with ultrapure water, methanol or another suitable
solvent (depending on the application) and wipe with a lint-free
cloth.

Rinse again with ultrapure water and dry with a lint-free cloth.

4 Reinserting the lens

Insert a new inner seal (12-4) into the measuring cell; make sure
that it lies flat and centered in the recess of the measuring cell.

Reinsert the lens (12-3) with its flat side facing inwards; make
sure that it lies flat and centered in the recess of the measuring
cell.

Insert a new outer seal into the retaining screw (12-2).

Reinsert the retaining screw (12-1) and tighten it with a slotted
screwdriver, size 5.

5 Cleaning the second lens

Repeat steps 2 to 4 on the opposite side of the cell holder.

6 Reinserting the flow-through cell

Follow steps 3 and 4 of "Installing the flow-through cell", page 16.