beautypg.com

8 technical specifications, 1 reference conditions, 2 uv/vis detector – Metrohm 944 Professional UV/VIS Detector Vario User Manual

Page 43: Reference conditions, Uv/vis detector

background image

■■■■■■■■■■■■■■■■■■■■■■

8 Technical specifications

944 Professional UV/VIS Detector Vario (2.944.0010)

■■■■■■■■

35

8 Technical specifications

8.1

Reference conditions

The technical specifications listed in this chapter refer to the following ref-
erence conditions:

Ambient tempera-
ture

+25 °C (± 3 °C)

Instrument status

> 30 min in operation with both lamps switched on.

8.2

UV/VIS detector

Detector type

Diode array
detector

512 diodes

Measuring range

Absorption

–2.0 - +2.0 AU

Number of
channels

8 measuring channels, 4 of them readable in analog mode

Resolution

2.5 x 10

–7

AU

Noise

2.5 x 10

–5

AU (at a 1/s data rate)

Drift

at reference
conditions

5 x 10

–4

AU/h

Wavelength range

Wavelength
range

λ

190 - 900 nm

Bandwidth

±1 - 50 nm

Accuracy, abso-
lute

±3 nm

Stability

±1 nm (across temperature range)

Optical resolu-
tion

5 nm (at 254 nm)

Measuring interval

Data rate for
each channel

0.5 - 50 samples/sec