Metrohm tiamo 2.4 Installation User Manual
Page 9

8.101.8056EN
9
2 If desired, export the methods in addition to the backup with the configu-
ration database (see above) into an external directory. In the Method
program part, select the File
▶ Method manager... menu item. Select
the desired Method group, select all methods to be exported and click
on [Edit]
▶ Export…. Select the desired export directory and click on
[OK]. The methods are exported into the desired directory. Afterwards,
copy the method files to CD/DVD.
3 Save all determination databases created by you into an external directory
using the backup function of tiamo. In the Database program part,
select the File
▶ Database manager… menu item. Select the desired
database and click on [Backup]. Select the Backup directory, enter a
name for the backup file and click on [Start]. The backup is created.
Afterwards, copy the backup file to CD/DVD.
4 Save further files created by you in the installation directory (e.g. export
files) into an external directory or to CD/DVD.
5 Close tiamo. Insert the installation CD and double-click on the setup.exe
file. Then select Repair/Update and click on [Next >].
6 When updating older tiamo versions (1.x) to tiamo 2.x, you have to enter
your update license code and then click on [Next >].