15 mounting the base plate, Figure 28, Mounting the base plate – Metrohm 855 Robotic Titrosampler User Manual
Page 42
3.15 Mounting the base plate
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
855 Robotic Titrosampler
Switch the barcode reader to programming mode and make the following
settings:
1
■
Select the keyboard layout for the desired country (USA, Germany,
France, Spain, Switzerland (German)). This setting must match the
setting in the control software.
■
Make sure that the Ctrl characters (ASCII 00 to 31) are allowed to
be sent.
■
Adjust the settings so that the ASCII character 02 (STX or Ctrl B) is
sent as the first character as "Preamble" or "Prefix Code".
■
Adjust the settings so that the ASCII character 04 (EOT or Ctrl D) is
sent as the last character as "Postamble" or "Record Suffix" or "Post-
fix Code".
■
Exit programming mode.
3.15
Mounting the base plate
If it is needed by an application that a determination is not carried out on
the sample rack but in an external measuring cell, a stand plate can be
mounted. It can be placed on the left or on the right of a tower of the 855
Robotic Titrosampler . Any accessories parts can be placed on the support
rod of the stand plate.
Mounting the base plate
6.2001.070
1
2
3
4
5
6
Figure 28
Mounting the base plate