Figure 16, Mounting the distributor – Metrohm 855 Robotic Titrosampler User Manual
Page 31
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
855 Robotic Titrosampler
■■■■■■■■
23
Note
The opening of the tubing may need to be widened with a sharp
object (e.g. with a Phillips screwdriver).
A piece of sandpaper may be used to get a better grip on the tubing.
Do not extend the tubing end before having slid the union nut onto
the tubing.
4 Mount the outlet tubing
■
Remove the union nut of the right-hand connector of the distributor
and guide it over the end of the 6.1812.000 PTFE tubing. Pull
the end of the tubing over the connection nipple of the distributor
and fasten in place with the union nut.
The tubing leads to the aspiration pump (Pump 2) and can be cut
to the correct length.
Mounting the distributor
6.1808.170
1
2
Figure 16
Mounting the distributor
Proceed as follows:
1 Remove a chain link
■
Remove the clip of the third chain link of the guide chain. Pry out
the clip with a screwdriver on both sides of the chain link, as shown
in the preceding illustration.
2 Insert the distributor
■
Apply strong pressure to insert the 6.1808.170 distributor (with
the tubing connected) into the open chain link.