10 column thermostat, 11 conductivity measuring system – Metrohm 850 Professional IC AnCat Nonsuppressed User Manual
Page 99
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
850 Professional IC–AnCat – non-suppressed
■■■■■■■■
89
7.10
Column thermostat
Type
Peltier technique thermostat for two intelligent separation columns
Adjustable tem-
perature range
0…+ 80 °C, in increments of 0.1 °C
Heating
Ambient temperature +50 °C
Cooling
Ambient temperature –20 °C
Temperature
reproducibility
± 0.2 °C
Stability
< 0.05 °C
Heating up time
< 30 minutes from 20 to 50 °C
Cooling time
< 40 minutes from 50 to 20 °C
7.11
Conductivity measuring system
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm single range
Noise
< 0.1 nS at 1 µS/cm
Drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 3 nS/cm typical without suppression
Conductivity
detector
Cell volume
0.8 µL
Cell constant
■
Individual calibration data saved in the detector
■
adjustable range: 13.0…21.0 /cm
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)
Cell tempera-
ture
20…50 °C in increments of 5 °C