Metrohm 850 Professional IC AnCat Nonsuppressed User Manual
Page 93
■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
850 Professional IC–AnCat – non-suppressed
■■■■■■■■
83
Problem
Cause
Remedy
Sample – Blockage in the
sample path.
Check the sample path.
Sample – Sample loop not
(completely) filled.
Prolong the sample transfer time.
Sample – Gas bubbles in
the sample.
Use the sample degasser (see Chapter 3.13,
page 43) (if available).
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see Chapter 5.5.2, page
64).
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
High pressure pump –
defective piston seals.
Replace piston seals (see Chapter 5.5.2, page
64).
Pulsation damper not con-
nected.
Connect the pulsation damper (see Chapter
3.12, page 41).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 3.12,
page 41).
Background conduc-
tivity too high
Incorrect eluent.
Change eluent (see Chapter 5.4.2.3, page 63).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than expec-
ted
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.12.4, page 80).