3.2 Initial installation
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
850 Professional IC – Anion MSM-HC – MCS
■
Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.
■
Place the second pump tubing in the second tubing cartridge.
■
Place both tubing cartridges into the peristaltic pump.
6 Connecting the MSM-HC
■
Connect the capillary labeled with H2SO4 with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.
■
Connect the capillary labeled with H20 with a 6.2744.180 pump
tubing connection and a 6.2744.070 short pressure screw to the
peristaltic pump on the outlet end of the pump tubing for the
rinsing solution.
■
Guide the two capillaries labeled with Waste into a waste con-
tainer and fasten it there.
7 Connecting the instrument
■
Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.20.1, page 65).
■
Connect instrument to mains supply (see Chapter 3.20, page
65).
8 Initial start-up
■
Switch on the PC and start MagIC Net.
■
Switch on the instrument.
■
Deaerate the high pressure pump.
■
Set the contact pressure of the peristaltic pump
■
Rinse instrument without column.
9 Installing guard and separation column
■
Remove the 6.2744.040 coupling between the column inlet capil-
lary and the eluent inlet capillary of the MSM-HC.
■
Connect guard column (optional) (see Chapter 3.21, page 66)
– Fasten the guard column to the end of the column inlet
capillary according to the specifications contained in the
leaflet accompanying the guard column.
– Rinse guard column.