5 Troubleshooting
850 Professional IC – Anion MSM-HC – MCS
■■■■■■■■
107
Problem
Cause
Remedy
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 69).
Conductivity detec-
tor is not recognized
in the software
No connection.
■
Check the cable connection (35-
■
Turn the instrument off and on again (after
15 seconds).
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Connections – dead vol-
ume in the system.
Check the connections (see Chapter 3.5, page
18) (use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).
Guard column – diminished
performance.
■
Replace guard column (see Chapter 3.21,
page 66).
Separation column – dead
volume at column head.
■
Install separation column in reverse flow
direction (if permitted by the information
sheet).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 69).
Greater rise in the
baseline
MSM-HC – reduced
capacity.
Regenerate the MSM-HC (see Chapter
4.11.3.1, page 92).
Precision problems -
significant scatter-
ing of the measured
values
Sample – Gas bubbles in
the sample.
Use sample degasser (see Chapter 3.13, page
44).
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 3.14.1, page 45).
Sample – Rinsing volume
too low.
Prolong the rinsing time (see Chapter 4.8,
page 89).
Injection valve – defective.
Request Metrohm Service.
MCS – vacuum too low.
■
Check connectors. If they are ok:
■
Contact the Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 4.15.4, page 102).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 69).