3 installation, 1 about this chapter, 2 initial installation – Metrohm 850 Professional IC Anion MSM-HC MCS User Manual
Page 22: About this chapter, Initial installation

3.1 About this chapter
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
850 Professional IC – Anion MSM-HC – MCS
3 Installation
3.1
About this chapter
The Installation chapter contains
■
this overview
■
a brief set of instructions for the initial installation of the 850 Professio-
nal IC – Anion MSM-HC – MCS (see Chapter 3.2, page 12). At each
step you will find cross-references to more detailed installation instruc-
tions for individual components, should you require such aids.
■
an installation diagram (see Chapter 3.3, page 15), showing a com-
pletely installed 850 Professional IC – Anion MSM-HC – MCS.
■
several chapters (see chapter 3.4, page 18 and following) with
detailed installation instructions for all components, including those
that are already installed at the time the instrument is delivered.
3.2
Initial installation
Note
A number of the capillaries is already connected at the time the instru-
ment is delivered.
The following work steps must still be carried out:
Installing 850 Professional IC – Anion MSM-HC – MCS
1 Setting up the instrument
2 Installations on the rear of the instrument
■
Remove handle and rollers (see Chapter 3.6.1, page 20).
■
Remove transport locking screws (see Chapter 3.6.3, page 25).
■
Place the detector in the instrument and connect it (see Chapter
3.6.2, page 23).
■
Connect the leak sensor (see Chapter 3.6.4, page 25).
■
Connect drainage tubings (see Chapter 3.6.5, page 26).