4 Start-up
894 Professional CVS
■■■■■■■■
53
4 Start-up
The 894 Professional CVS is operated exclusively via the viva PC software.
You can find information on operating viva in the online help and in the
Tutorial CVS.
Proceed as follows for the initial start-up of the 894 Professional CVS:
Preparing the system for starting up
WARNING
Uncontrolled splashing of reagents
Splashing reagents may result in injuries.
Operate the 894 Professional CVS only with the measuring head in
place and the measuring head arm lowered.
WARNING
Drive belt in operation
Hair may become entangled or caught in the running drive belt, for
example.
Operate the 894 Professional CVS only with the measuring head cover
in place.
1 Insert the measuring head as described in "Preparing the measuring
2 Equip the measuring head with the electrodes (see Chapter 3.2.1,
3 Establish the tubing connections in the measuring head and those
between the 894 Professional CVS and the peripheral devices (see
Chapter 3.3, page 27).
4 Attach the measuring head cover and the stopper (3-5) to the mea-
suring head as described in "Installing the measuring head cover",
page 34.