Figure 28 – Metrohm 894 Professional CVS User Manual
Page 47

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
894 Professional CVS
■■■■■■■■
37
Connecting an FEP tubing to the waste canister
A 5-way tubing connector (6.1828.020) is required to connect FEP tubings
to a waste canister. Figure 43, page 69, provides an overview of the
tubing setup between the waste canister, 843 Pump Station and 894 Pro-
fessional CVS.
1
2
Figure 28
Connecting the 5-way tubing connector to a waste canister
Proceed as follows:
1 Unscrew the cap with the hole in the center from the waste canister.
2 Feed each of the five tubings from the 5-way tubing connector with
the couplings already installed from below through the hole of the
unscrewed cap.
3 Screw the cap with the 5-way tubing connector inserted back onto
the waste canister.
4 Screw the tubing nipples of the FEP tubings (e.g. 6.1805.530) to the
couplings of the tubings.
NOTE
In order to ensure that liquid is transported smoothly from and to the
canisters, the canisters must not be air-tight. If necessary, loosen the
screw cap a little.
Connecting an FEP tubing to a dosing unit with VMS
Proceed as follows:
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)