3.13 Sample degasser
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
881 Compact IC pro – Anion – MCS
1
2
3
4
Figure 19
Sample degasser
1
Sample degasser input
2
Sample degasser output
3
PEEK pressure screw, long
(6.2744.090)
4
Connection capillaries (6.1803.040)
Connecting the sample degasser
1 Remove and keep the threaded stoppers (6.2744.220) from the input
and output of the sample degasser.
2 Connect the end of the sample aspirating capillary (6.1803.040) con-
nected to the injection valve to the output of the sample degasser
(19-2), using a long PEEK pressure screw (19-3).
3 Connect the connection capillary (6.1803.040) to the input of the
sample degasser (19-1), using a long PEEK pressure screw (19-3).
4 Guide the other end of the connection capillary out of the instrument
through a capillary feed-through and connect it with the Sample Pro-
cessor, if applicable.
Caution
If the sample degasser is not used, the input and output must be
sealed with the threaded stoppers (6.2744.220).
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)