3 Installation
930 Compact IC Flex ChS/PP (2.930.1300)
■■■■■■■■
49
You can find the separation columns and guard columns currently avail-
able from Metrohm in the Metrohm IC Column Program or on the Inter-
in the Ion Chromatography product
area. A test chromatogram and a leaflet are provided along with each col-
umn. You can find detailed information on special IC applications in the
corresponding "Application Bulletins" or "Application Notes", which
are available on the Internet at
tions area or free of charge upon request from your responsible Metrohm
representative.
CAUTION
New separation columns are filled with solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow information from
the manufacturer).
NOTE
The separation column may not be connected until after the instrument
has already been put into operation once (see Chapter 3.15, page 44).
The guard column and separation column have to be replaced by a
coupling (6.2744.040) until then.