7 injection valve, 8 suppressor, 9 peristaltic pump – Metrohm 930 Compact IC Flex ChS/PP User Manual
Page 104: Injection valve, Suppressor, Peristaltic pump

7.7 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
96
■■■■■■■■
930 Compact IC Flex ChS/PP (2.930.1300)
Safety shutdown
Function
Automatic shutdown upon reaching the pressure limit values
Maximum pres-
sure limit
■
Adjustable from 0.1 - 50 MPa (1 - 500 bar)
■
The pump is automatically shut down at the first piston stroke
above the maximum limit value
Minimum pres-
sure limit
■
Adjustable from 0 - 49 MPa (0 - 490 bar)
■
The shutdown mechanism is inactive at 0 MPa
■
The shutdown mechanism becomes active 2 minutes after system
start
■
The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit
7.7
Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
7.8
Suppressor
Resistance to sol-
vents
No restriction
Switching time
typ. 100 ms
7.9
Peristaltic pump
Type
2-channel peristaltic pump
Direction of rota-
tion
Clockwise/counterclockwise
Rotational speed
0 - 42 rpm in 7 levels of 6 rpm each
Pumping charac-
teristics
0.3 mL/min at 18 rpm; with standard pump tubing (6.1826.420)
Pump tubing
material
Recommended: PharMed® (Ismaprene)