4. Appendix
700 Dosino, Instructions for Use
39
(
)
1
2
−
−
=
∑
n
V
V
s
i
s : repeatability standard deviation
V
i
: distributed volume
n : number of distribution
Please note
•
The above-mentioned limits are related to water at 20 °C and
are valid for dosing units with glass cylinders only.
•
The nominal volume of the dosing cylinders are printed on the
housing of the dosing units.
•
The above-mentioned limits are valid for the useful volume ran-
ge of 10 % to 100 % of the nominal volume.
•
The error limits relate to the ISO 8655-3 Draft International
Standard (Piston-operated volumetric apparatus - part 3: Piston
burettes), which is to come to effect in the year 2000.
•
Metrohm burets with glass cylinders in general have a better
quality performance (systematic and random error) than the
maximum error limits implied. Check your dosing system ac-
cording to the ISO test specification.
Possible sources of dosing errors
Influencing parameters are:
•
Air bubbles in the dosing cylinder or tubings.
•
Leaky piston/cylinder system.
•
Leaky tubing connections.
•
Differences in hydrostatic pressure of the liquid level in the
supply flask versus the dosing tip.
•
Viscosity and/or flow characteristics of the liquid to be dosed.